Fluke 434 User Manual Page 108

  • Download
  • Add to my manuals
  • Print
  • Page
    / 122
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 107
Fluke 433/434
Users Manual
21-6
Measurements Measurement Range Resolution Accuracy
RMS voltage in Transient
Function
10 ... 1000 Vrms 1 Vrms ± 2.5 % of Vnom
Transient voltage
(TTRANS > 10 µs)
0 ... ± 6000 Vpeak 1 V ± 15 % of meas
Flicker
Settings Adjustment Range Resolution Additional
Information
Steady Voltage
Tolerance
0.0 ... 10.0 % of Vnom 0.1 %
Minimum Steady Time 0.0 ... 10.0 s 0.1 s
Maximum Deviation
Threshold
0.0 ... 10.0 % of Vnom 0.1 %
Measurement Measurement Range Resolution Accuracy
PF5 0.00 ... 20.00 0.01
P1min 0.00 ... 20.00 0.01
Pst 0.00 ... 20.00 0.01
Plt 0.00 ... 20.00 0.01
Within ±5 % of tabulated
values according
IEC61000-4-15
Dc 0.0 ... ± 100.0 % 0.1 % ± 1 % (if Umeas Vnom)
DMAX 0.0 ... ± 100.0 % 0.1 % ± 1 % (if Umeas Vnom)
TDEX 0.000 ... 9.999 s 10 ms 20 ms (at F = 50 Hz)
CURRENT MEASUREMENTS
Current Inputs
Item Specification Additional Information
Nominal input Range 0 - ±5.625 Vpeak 0 - 3.97 Vrms sinewave
Current Clamp Sensitivity 0.1 , 1, 10, 100, 1000 mV/A
variable:
Input impedance 50 k.ohm
Bandwidth >10 kHz
Voltage resolution 1 mV
Page view 107
1 2 ... 103 104 105 106 107 108 109 110 111 112 113 ... 121 122

Comments to this Manuals

No comments