Data Menu 5
5-19
any hour of the day and automatically reset the statistics on any hour. Measurement
statistics are kept as measurements are made. The current statistics can be displayed in
any STATS type zone. The statistics can also be stored automatically at the end of the
day and reset automatically at the beginning of the next day. The DAILY STATS menu
options are STATS SETTING, STATS VIEW,
STATS PRINT, STATS RESET, and STATS CLEAR. The DAILY STATS can be
password-protected.
avu025.jpg
Figure 5-24. Daily Stats
The A D buttons are used to select the menu option. The B C buttons are used to change
the setting parameters. Press the E button to save the new setting. Press the F button
to cancel any changes and exit to the menu.
Stats Setting
The STATS SETTING function sets the daily statistics recording options. The AUT REC
enables (ON) or disables (OFF) automatic recording of the statistics. The REC TIM may
be set from 00:00 to 23:00. The REC TIM function sets the hour just prior to which the
statistics are automatically recorded when AUT REC is ON. If REC TIM is 00:00, the
statistics are recorded after the last measurement of the day.
AUT RES enables (ON) or disables (OFF) automatic reset of the statistics and may be set
from 00:00 to 23:00. RES TIM sets the hour when the statistics are automatically reset if
AUT RES is ON. If RES TIM is 00:00, the statistics are reset just prior to the first
measurement of the day.
For example, to automatically store statistics for each day for measurements made in a
laboratory only between 8:00 in the morning and 5:00 (17:00) in the afternoon, set the
AUT REC to ON, REC TIM to 17:00, AUT RES to ON, and RES TIM to 08:00. Note
that statistics can also be manually reset using the STATS RESET function. DO NOT
manually reset the statistics if automatic recording is desired. The DAILY STATS
settings and STATS RESET can be password-protected to prevent inadvertent resetting
of the statistics. DAILY STATS cannot be manually stored.