Fluke 5020A User Manual Page 51

  • Download
  • Add to my manuals
  • Print
  • Page
    / 142
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 50
Menu Functions
Data Menu 5
5-15
Caution
When the memory has been filled, the oldest block of data will
be discarded to allow new data to be recorded.
Measurement storage capacity can range between 28 hours and 2 years depending on the
record period and enabled channels. Refer to Table 5-2, Approximate Data Storage
Capacity.
Table 5-2. Aproximate Data Storage Capacity
Record Period Storage, 1 channel Storage, 2 channels
1 second 56 hours 28 hours
10 seconds 20 days 10 days
1 minute 18 weeks 9 weeks
5 minutes 20 months 10 months
10 minutes 24 months 20 months
30 minutes 24 months 24 months
1 hour 24 months 24 months
avu019.jpg
Figure 5-18. Record Setting
Use the A D buttons to select the setting to change. The B C buttons are used to change
the setting parameters. Press the E button to save the new setting. Press the F button
to cancel any changes and exit to the menu.
Data View
The DATA VIEW function allows viewing recorded data in NUMERIC or GRAPHIC
format.
Page view 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 141 142

Comments to this Manuals

No comments