Maintenance
PERFORMANCE TEST
6
6-7
6-5. AC Voltage Test (Option -09 Only)
The following procedure may be used to verify the accuracy of the VAC function:
1. Ensure the 8842A is on and warmed up for at least 1 hour.
2. Select the VAC function and the slow (S) reading rate.
3. Connect the AC Calibrator to provide a voltage input to the HI and LO INPUT
terminals.
4. (Low- and Mid-Frequency Test.) For each step in Table 6-3, select the indicated
range, set the AC Calibrator for the specified input, and verify that the displayed
reading is within the limits shown for each reading rate.
NOTE
This procedure tests the extremes of each range. You may shorten the
procedure by testing only the "quick test points" indicated in Table 6-3 with
asterisks.
5. (High-Frequency Test.) For each step in Table 6-4, select the indicated range, set the
AC Calibrator for the specified input, and verify that the displayed reading is within
the limits shown for each reading rate.
NOTE
This procedure tests the extremes of each range. You may shorten the
procedure by testing only the "quick test points" indicated in Table 6-4 with
asterisks.
6. Set the AC calibrator to standby and disconnect it from the 8842A.
Table 6-3. Low- and Mid-Frequency AC Voltage Test
Input Display Readings
Step
Number
Range
Voltage Frequency
Error in
Counts
Minimum Maximum
1 2V 0.01000V 200 Hz 181 0.00819 VAC 0.01181 VAC
2 2V 0.10000V 200 Hz 87 .09913 VAC .10087 VAC
3* 2V 0.30000V 200 Hz 101 0.29899 VAC 0.30101 VAC
4 2V 1.00000V 200 Hz 150 0.99850 VAC 1.00150 VAC
5* 2V 1.90000V 200 Hz 213 1.89787 VAC 1.90213 VAC
6 2V 0.10000V 20 Hz 220 0.09780 VAC 0.10220 VAC
7 2V 1.90000V 20 Hz 2380 1.87620 VAC 1.92380 VAC
8 2V .10000V 45 Hz 135 0.09865 VAC 0.10135 VAC
9 2V 1.90000V 45 Hz 765 1.89235 VAC 1.90765 VAC
10* 200 mV 0.001000V 100 Hz 204 00.796 mVAC 190.252 mVAC
11 200 mV 0.190000V 20 kHz 252 189.748 mVAC 190.252 MvAC
* Quick Test points