50S & 50D
Thermometer
Calibration Information
WWarning
These service instructions are for use by qualified personnel only. To avoid
electric shock, do not perform any servicing unless you are qualified to do so.
Introduction
This document contains specifications, general maintenance information, performance tests, calibration
procedures, and a list of user replaceable parts for the Fluke 50S and 50D Thermometers, hereafter referred
to as “the thermometer”.
Service
To contact Fluke, order parts, or locate a Fluke service center, call one of the following telephone numbers:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com.
Warranty
This Fluke product will be free from defects in material and workmanship for one year from the date of
purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse
or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty
on Fluke’s behalf. To obtain service during the warranty period, send your defective tester to the nearest
Fluke Authorized Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR
A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation
of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to
you.
Fluke Corporation
P.O. Box 9090
Everett WA
98206-9090
Fluke Europe B.V.
P.O. Box 1186
5602 B.D.
Eindhoven
The Netherlands
Page view 0
1 2 3 4 5 6 ... 13 14

Summary of Contents

Page 1 - 50S & 50D

PN 1281009October 1999 Rev.1, 3/03©1999-2003 Fluke Corporation. All rights reserved. Printed in U.S.A.All product names are trademarks of their respec

Page 2 - International Symbols

50S & 50D ThermometerCalibration Information10position until the display is locked in self-test (all LCD segments on). This thermometer is now in

Page 3 - Basic Maintenance

Calibration11FLUKE 52-1601zf09f.epsFigure 6. A1 Main PCA AdjustmentsOptimizing Thermometer Performance in the Negative Temperature RangePerform the fo

Page 4 - Disassembly/Reassembly

50S & 50D ThermometerCalibration Information125. Allow the system to stabilize, then adjust T1 or T2 offset adjustments for a temperature display

Page 5 - WWarning

Replaceable Parts13Replaceable PartsRefer to Table 4 for all replacement part numbers and Figure 8 for part locations. Unless specified in the“Model N

Page 6 - Performance Tests

50S & 50D ThermometerCalibration Information141561011271331181692X4X4X4X245zf08f.epsFigure 8. Exploded View Diagram (Model 50D is Shown)

Page 7 - MAXMIN HOLD REC

50S & 50D ThermometerCalibration Information2International SymbolsTable 1 identifies the international electrical symbols used in this document.Ta

Page 8 - Ice Bath Test

Specifications3SpecificationsGeneralWeight 280 g (10 oz)Dimensions(without holster)2.8 cm × 7.5 cm × 16.6 cm(1.1 in × 3 in × 6.6 in)Battery Standard 9

Page 9

50S & 50D ThermometerCalibration Information4Clean the thermometer case with a mild detergent and water.The pca may be washed with isopropyl alcoh

Page 10 - 50S & 50D Thermometer

Basic Maintenance5Changing the BatteryWWarning• Replace the battery as soon as the low battery indicator (N) appears. Thepossibility of false reading

Page 11 - Calibration

50S & 50D ThermometerCalibration Information6Required Test EquipmentTable 2 lists the required test equipment needed to perform the performance te

Page 12

Performance Tests7KJMAXMIN HOLD RECCF50S Display 50D Display- +T1 T2 T1-T2KJHOLDCF- +zf03f.epsFigure 3. Display SegmentsWide Range Performance Test Pr

Page 13 - Replaceable Parts

50S & 50D ThermometerCalibration Information85520AIce bathThermocouple wire50S or 50DThermometer +_Copper wirezf05f.epsFigure 4. Ice

Page 14

Calibration9Table 3. Performance Test ValuesThermocouple Input Voltage Display ReadingType (mV DC)oCoFK -5.587 -182.0 ±(0.9) -295.6 ±(1.6)K -3.211 -89