Specifications
25
Specifications
Feature 561 566 568
IR Temperature Range
-40 °C to 550 °C
(-40 °F to 1022 °F)
-40 °C to 650 °C
(-40 °F to 1202 °F)
-40 °C to 800 °C
(-40 °F to 1472 °F)
Accuracy
<0 °C: ±(1.0 °C + 0.1 °/1 °C)
>0 °C: ±1 % or ± 1.0 °C, whichever is greater
(<32 F±2 °F ±0.1 °/1 °F)
>32 °F: ±1 % or ±2 °F), whichever is greater
Repeatability
±0.5 % of reading or
±1 °C (2 °F),
whichever is greater.
±0.5 % of reading or ±0.5 °C (1 °F),
whichever is greater.
Display Resolution 0.1 °C / 0.1 °F
Spectral Response 8 to 14 μm
Response Time (95 %) <500 ms
K-Type Thermocouple
Input Temperature Range
0 °C to 100 °C
(32 °F to 212 °F)
-270 °C to 1372 °C (-454 °F to 2501 °F)
K-Type Thermocouple
Input Accuracy
Input accuracy ±2.2 °C
(±4 °F)
-270 °C to -40 °C: ±(1 °C + 0.2 °/1 °C)
(-454 °F to -40 °F: ±(2 °F + 0.2 °/1 °F))
-40 °C to 1372 °C: ±1 % or 1 °C (-40 °F to
2501 °F: ±1 % or 2 °F), whichever is greater
K-Type Thermocouple
Resolution
0.1 °C (0.1 °F) 0.1 °C/0.1 °F
Distance:Spot
(90 % energy)
12:1 30:1 50:1
Laser sighting Single laser, output <1 mW Class II, wavelength 630 to 670 nm
Emissivity Lo, Med, Hi
Digitally adjustable from 0.10 to 1.00 by 0.01
or via built-in table of common materials
Data storage - 20 points 99 points
Communication none USB 2.0
Operating Altitude 3000 meters above mean sea level
Storage Altitude 12,000 meters above mean sea level
Relative Humidity 10 % to 90 % RH non-condensing up to 30 °C (86 °F)
Operating Temperature
0 °C to 50 °C
(32 °F to 122 °F)
0 °C to 50 °C (32 °F to 122 °F)
Storage Temperature
-20 °C to 65 °C
(-4 °F to 149 °F)
-20 °C to 60 °C (-4 °F to 149 °F)
Vibration 2.5 G, IEC 68-2-6
Weight 0.322 kg (0.7099 lb)
Dimensions 17.69 cm (6.965 in) H x 16.36 cm (6.441 in) L x 5.18 cm (2.039 in) W
Power
2 AA /LR6 Batteries (alkaline or NiCD)
2 AA /LR6 Batteries
or USB connection
when used with a
PC
Battery Life
12 hours with laser and backlight on; 100 hours with laser and backlight
off, at 100 % duty cycle (thermometer continuously on)
CE Certification
EN/IEC 61326-1:2006, Class B, Criteria A