Fluke 1529 Chub-E4 User Manual Page 131

  • Download
  • Add to my manuals
  • Print
  • Page
    / 154
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 130
bit 4: measurement invalid
bits 5-15: not used
8.4.12.18 STATus:QUEStionable:CONDition?
Returns the Questionable Status Condition Register.
STAT:QUES:COND?
16
The assignment of the bits in this register is as follows:
bits 0-3: not used
bit 4: measurement invalid
bits 5-15: not used
8.4.12.19 STATus:QUEStionable:ENABle? [MIN|MAX|DEF]
Returns the Questionable Status Enable Register.
STAT:QUES:ENAB?
255
8.4.12.20 STATus:QUEStionable:ENABle <num>|MIN|MAX|DEF
Sets the Questionable Status Enable Register.
STAT:QUES:ENAB 10
125
8 Digital Communications Interface
Commands
Page view 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 153 154