Fluke PM-3370B User Manual Page 137

  • Download
  • Add to my manuals
  • Print
  • Page
    / 297
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 136
COMMAND REFERENCE 4 - 37
CALCulate<n>:MATH:STATe
Syntax:
CALCulate<n>:MATH:STATe <Boolean>
<n> [1] | 2
Query form:
CALCulate<n>:MATH:STATe?
Response:
0 | 1
0 Mathematics function turned off.
1 Mathematics function turned on.
Description:
This command switches the specified mathematics function on or off. If the
mathematics function is switched on, the internal scale and offset are reset to
initial values. The result of the mathematics function is stored in M1_1 for
CALCulate1 and in M2_1 for CALCulate2.
After a
*
RST command, the mathematics function is turned off.
Example:
Send
CALCulate:MATH (CH1-CH2) Selects MATH1 channel 1 - 2.
Send
CALCulate:MATH:STATe ON Switches MATH1 function on.
Front panel compliance:
The CALCulate1 and CALCulate2 commands use the MATH1 and MATH2
features of the CombiScope instrument.
Page view 136