Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL TECHNICAL INFORMATION When comparing two products features and specification us
ACV & DCV 4 wire input. Selectable 2 and 4-wire measurement. Unavailable 4-wire sense measurement support improved accuracy when calibrating cal
8508A vs 3458/HFL Ease of Use Parameters Comments Rear input switched remotely using IEEE 488 commands Yes No The 3458/HFL must be switched manu
8508A Options and Accessories: Model Description8508A 8.5 digit Reference Multimeter 8508A/01 8.5 digit Reference Multimeter with Rear Input Rat
Notes:
DC Voltage ±(ppm of setting + ppm of range), relative to calibration standards used.Calibration Interval 365 day, ±1°C of last TCALRange 1281 HP3
DC Current ±(ppm of setting + ppm of range), relative to calibration standards used.Calibration Interval 365 day, ±1°C of last TCALRange 1281 HP34
Resistance ±(ppm of setting + ppm of range), relative to calibration standards used.Calibration Interval 365 day, ±1°C of last TCALRange 1281 HP3458/
AC Current ±(ppm of setting + ppm of range), relative to calibration standards used.Calibration Interval 365 day, ±1°C of last TCALRange Frequency 12
AC Voltage ±(ppm of setting + ppm of range), relative to calibration standards used.Calibration Interval 365 day, ±1°C of last TCALRange Frequency 12
AC Voltage ±(ppm of setting + ppm of range), relative to calibration standards used.Calibration Interval 365 day, ±1°C of last TCALRange Frequency 1
Fluke 8508A 3458/HFL 8508A vs 3458/HFL Functional & Versatility Parameters Comments Full Scale (dc & ac Voltage)
8508A vs 3458/HFL Functional & Versatility Parameters Comments High Resistance Measurement (compliance voltage) Test Voltage or Full Scale stim