PN 1554518 January 2000 ©2000 Fluke Corporation. All rights reserved. Printed in U.S.A. 1
Model 65
Infrared Thermometer
Calibration Information
Introduction
This calibration information for the Fluke 65 Infrared Thermometer (also referred to as “the UUT”) provides
the following information:
Safety information
Specifications
Basic maintenance
Equipment required for performance tests and calibration
Performance test procedures
Calibration adjustment procedures
List of replaceable parts
For complete operating instructions, refer to the Model 65 Infrared Thermometer Instruction Sheet.
Contacting Fluke
To locate an authorized service center, visit us on the World Wide Web at www.fluke.com. Or, call Fluke as
follows:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Page view 0
1 2 3 4 5 6 7 8

Summary of Contents

Page 1 - Model 65

®PN 1554518 January 2000 ©2000 Fluke Corporation. All rights reserved. Printed in U.S.A. 1Model 65Infrared ThermometerCalibration InformationIntrodu

Page 2 - Safety Information

Model 65Calibration Information2Safety InformationA WarningTo avoid possible personal injury:• The information provided in this manual is for the use

Page 3 - Specifications

Specifications3SpecificationsMeasurement Range −40 °C to +500 °C (−40 °F to +932 °F)Operating Temperature (Ambient) 0 °C to +50 °C (+32 °F to +122 °F)

Page 4 - Maintenance

Model 65Calibration Information4MaintenanceInstalling the BatteriesRemove the holster to access the battery door. Figure 1 shows how to replace the ba

Page 5 - Performance Tests

Required Equipment5Required EquipmentTable 1 shows the equipment required for the performance tests and calibration adjustments.Table 1. Required Equi

Page 6 - Calibration Adjustments

Model 65Calibration Information6Table 2. Performance Test ValuesTemperature at 0.95 Emissivity UUT Reading Limits1, 2Room temperature Room temperature

Page 7

Calibration Adjustments7MINMAX°C°FRoom temperaturecalibration modeHigh temperaturecalibration modeEnd calibrationmodePress to exitcalibration mode

Page 8 - Replacement Parts

Model 65Calibration Information88. Press D to enter the high-temperature adjustment mode (display shows 2_H), as shown in Figure 2.9. Press G. The d