9933 TableW areUser’ s GuideRev. 592101Hart Scientific
3. When prompted, enter the product serial number. The serial number isprovided on a sticker typically located on the CD pouch on the insideback cover
The TableWare main display screen (Figure 1) displays the software menu andtoolbar.Enabled and disabledAll menu items and toolbar icons are enabled.Sc
The following icons are available on the toolbar:Defaults - Setup default settingsPlatinum probe coefficient calculation - Calculate platinum probecoe
2 File menuThe File menu (Figure 2) provides options for setting program defaults, select-ing the default printer, and exiting the software.Enabled an
Enabled and disabledAll tabs are enabled at all times.Dialog informationThe TableWare Defaults dialog allows the user to set default values or parame-
The “Show help balloons” check box toggles the display of the help balloons.The default is selected which means the yellow help balloons are displayed
2.1.2 Defaults - TablesThe Tables tab (Figure 5) is displayed when the File|Defaults menu option isselected and the tab labeled Tables is selected.Ena
2.1.3 Defaults - ResolutionsThe Resolutions tab (Figure 6) is displayed when the File|Defaults menu optionis selected and the tab labeled Resolutions
2.2 Setup printerThe setup printer menu option allows the user to select and configure the de-fault printer. For TableWare, the orientation should alw
3 Coefficients menuThe Coefficients Menu (Figure 7) provides options for calculating coefficientsfor Platinum, Thermistor, or Thermocouple probes.Sele
Rev. 592101Fluke CorporationHart Scientific Division799 E. Utah Valley Drive American Fork, UT 84003-9775 USAPhone: +1.801.763.1600Telefax: +1.801.7
Dialog informationThe Probe Information dialog allows the user to enter two captions and valuesto identify the test probe or UUT (Unit Under Test) for
default is Range 4. The choices for high range are Range 6 through Range11, or None. The default is Range 8.•If the type selected is Polynomial, the o
Note: The accepted thermistor model is based upon the logarithmic resis-tance-temperature characteristic in terms of absolute temperature (Kel-vin). T
Dialog informationThe type of coefficient choices are: B, E, J, K, N, R, S, T, and Gold-Platinum.The default is Type K.The temperature range is depend
Enabled and disabledThe Calculate button is not enabled until the required data is entered and thescales are selected.Dialog informationThe user must
Callendar-Van Dusen (CVD), or IPTS-68 is selected as the type of coefficientsto calculate or from the Thermistor Probe Coefficients dialog.Enabled and
3.8, Methods Used for Calculating Coefficients, contains additionalinformation.The Import button reads raw data in from an ASCII text file. The data m
3.3.3 Thermocouple coefficientsThis Raw Data dialog (Figure 15) is displayed when the Continue button is se-lected from the Thermocouple Probe Coeffic
Note: If CJC readings are entered in temperature, the standard thermo-couple coefficients for the selected thermocouple type are used to calcu-late th
Note: The International Settings must be set according to the informationin the Installation section before attempting to import or export raw data.Fo
Table of Contents1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 What is TableWare?. . . .
Example: ASCII text file for ITS-90 (both low and high range) coefficients.0.010100.025–100.00,-38.84359.2564,83.2587156.6056,232.0110,419.4314,660.43
3.4.2 Polynomial, CVD, and IPTS-68 coefficientsThe raw data for calculating Polynomial, CVD and IPTS-68 coefficients can beimported from an ASCII text
Example: ASCII text file for IPTS-68 coefficients.–100.00,0.000,232.0110,419.431459.2564,100.025,188.6534,256.1023Note: If a specific set of temperatu
If the third line does not exist, this software does not compensate the voltagereadings.Example: ASCII text file for Thermocouple coefficients.–100.0,
Dialog informationThe calculated coefficients are displayed in the Coefficients list. Use the scrollbar, if necessary, to view the coefficients. The s
should always fit on one letter-size (A4) sheet of paper. Select the OK button toprint the report.The Table Options button displays the Table Options
ton is selected on the Coefficients and Residuals dialog or when the OK buttonis selected on the Enter Coefficients dialog.Enabled and disabledControl
selecting either the Export or Print Table button.Enter the date exactly as it is toappear on the report. No formatting or validation is performed on
Dialog informationThe Type and Range tab allows the user to select the type of table to generateand the temperature range for the table. The choices f
Enabled and disabledThe resolution selection boxes are disabled unless the “Use custom resolutions”option is selected.Dialog informationThe Resolution
3.5.1.3 Table options—other settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363.6 Table temperature r
3.5.1.3 Table options—other settingsThe Other Settings tab is displayed after selecting the Other Settings tab on theTable Options dialog.Enabled and
efficients used. See Table 1 on page 37 for details on the temperature range forspecific coefficients.373 Coefficients menuTable temperature range cha
3.7 Requirements for calculating coefficientsCertain requirements apply depending on the type of coefficients beingcalculated.3.7.1 ITS-90 coefficient
b) All Ranges require the Triple Point of Water set-point and thefollowing:i) Range 6 requires at least 4 and no more than 10 set-points.ii) Range 7 r
a) One set-point must be within a few degrees of 0°C (32°F or273.15K) (preferably as close to 0°C as possible).b) One set-point must be below 0°C.c) T
c) DELTAd) BETA (only calculated if criteria explained above is met)6. Residuals are calculated for each set-point used in calculating the coeffi-cien
a) T(R)—Temperature as a function of Resistance: A, B, ... N (depend-ing on the order of the polynomial)b) R(T)—Resistance as a function of Temperatur
e) Type Nf) Type Rg) Type Sh) Type Ti) Type AuPt (Gold-Platinum)5. The temperature range for the Thermocouple type must be selected. Thefollowing rang
b) c29. The following standard Thermocouple coefficients are also producedunaltered:TypeRange 1 Coefficients Range 2 Coefficients Range 3 Coefficients
solve for two coefficients. The example below shows a set of four equationsrepresenting data at five temperatures ( Rtpwand resistance at four test te
iiiFiguresFigure 1 TableWare Main Display Screen . . . . . . . . . . . . . . . . . . . . . 6Figure 2 File Menu . . . . . . .
The software allows for over determined solutions. Thermocouple probes re-quire polynomials expressed in terms of μV vs. temperature.3.9 Software outp
General tab. Other settings on the General tab can be changed to manipulate thedefault behavior of the Print Preview window.Enabled and disabledThe en
Print - Prints all pages of the document in the Print Preview windowFirst page - Navigates to the first page of the document (if more thanone page)Pre
493 Coefficients menuPrint preview windowFigure 27 Sample report or exported ASCII text file
9933 TableWareUser’s Guide50Figure 28 Sample table
4 Table menuThe Table menu (Figure 29) provides options for generating Temperature vs.Resistance, Temperature vs. Ratio and Temperature vs. EMF tables
Dialog informationEnter the coefficients to be used to generate the table. A value must be enteredinto each of the boxes provided.•For ITS-90, RTPW is
5 Help MenuThe Help Menu (Figure 31) provides a method for accessing help through thehelp file contents, search on a specific topic of the software, a
5.4 Technical supportThe Technical Support menu option displays the Technical Support topic in thehelp file.Please contact one of the following author
Fluke South East Asia Pte Ltd.Fluke ASEAN Regional OfficeService Center60 Alexandra Terrace #03-16The Comtech (Lobby D)118502SINGAPOREPhone: +65 6799-
1 Introduction1.1 What is TableWare?TableWare is designed to calculate characterization coefficients for platinum,thermistor, and thermocouple probes.
software version number, the year the version was released, and specific infor-mation about the computer are displayed on this dialog.9933 TableWareUs
6 Messages6.1 TableWareMessagesMessage “The [report][table] was successfully exported to filename.”Message “The raw data that was entered may contain
Section 3.6, Table Temperature Range Chart. If the selected minimum andmaximum temperatures are too extreme, the algorithms used to calculate the ta-b
Message “An error occurred while calculating coefficient type coefficients.Make sure the CALCCOEF.DLL file is located in the \WINDOWS\SYSTEMor \WINNT\
7 Glossarycoefficients - Values used to define characteristics of a probe.help balloons - The yellow boxes with words that appear below and to the sid
and/or Maximum Temp. setting fell outside the recommended tempera-ture range in °C.5. In version 2.2, an Overflow error sometimes occurred when attemp
If you want to create a document that includes the report and table, follow thesteps below:a. Set the starting page for the table to “2”.b. Print the
be used) the reports and tables are printed to, or which printer driver isused.5. The date that prints on the report and table can now be set by the u
Comments to this Manuals